Table 1c: Accuracy to YSI within CGM Glucose Ranges (Pediatric*; n=129)
CGM
Glucose
Level
(mg/dL)
Number
of CGM-
Reference
Pairs
Percent
Within
±15 mg/dL
Percent
Within
±20 mg/dL
Percent
Within
±40 mg/dL
Percent
Within
±15%
Percent
Within
±20%
Percent
Within
±40%
Mean
bias
(mg/dL)
MARD
(%)
<54 139 71.9 79.1 97.1 -9.9 17.1
54-69 863 86.4 90.5 97.1 -4.9 12.0
70-180 2690 77.4 87.6 98.7 -3.4 10.6
181-250 1236 86.0 94.7 99.7 -8.9 8.3
>250 1618 92.2 97.7 99.8 -2.2 7.2
*
Includes children 6-17 years of age. No YSI measurements were obtained for children 4-5 years of age.
System range is 40-400 mg/dL.
189