Concurrence of System and Reference (CGM vs. YSI)
The percentage of concurring glucose values (CGM vs. YSI) in each glucose reference range is
presented for each CGM range in Tables 3a and 3b and for each YSI range in Tables 3c and 3d.
For example, in the Adult study, when the System glucose readings were within the 81 to
120 mg/dL range, actual blood glucose values were between 40 and 60 mg/dL 0.2% of the time,
between 61 and 80 mg/dL 11.0% of the time, between 81 and 120 mg/dL 70.1% of the time,
between 121 and 160 mg/dL 17.8% of the time, between 161 and 200 mg/dL 0.8% of the time, and
between 201 and 250 mg/dL 0.1% of the time.
194