Glucose Reading Availability
The System is designed to show a Sensor glucose reading after each scan that is performed
throughout the wear period after the start-up time. Tables 8a and 8b show the glucose reading
capture rate for each day of the wear duration.
Table 8a: Glucose Reading Capture Rate Over Wear Duration (Adult; n=146)
Day of Wear Number of Sensors Capture Rate (%)
1 146 98.3
2 145 98.1
3 143 98.3
4 140 98.3
5 138 98.4
6 135 98.3
7 134 98.4
8 131 98.4
9 128 98.4
10 123 98.4
11 120 98.4
12 113 98.5
13 112 98.5
14 104 98.6