Precision
Precision of the System was evaluated by comparing the results from two separate Sensors worn on
the same subject at the same time. Table 9a provides data from 146 subjects in the Adult study;
Table 9b provides data from 137 subjects in the Pediatric study. For adults, the paired absolute
relative dierence (PARD) between the two Sensors was 8.1% with coecient of variation (CV) of
5.7%. For children ages 4-5, PARD was 6.7% with CV of 4.8%. For children ages 6-17, PARD was 8.2%
with CV of 5.8%. Paired absolute dierence (PAD) is a measurement of absolute dierence (in mg/dL)
between paired CGM readings, while PARD is the absolute relative dierence (in %) between paired
CGM readings.
Table 9a: Overall between Sensor Precision (Adult; n=146)
Coecient of
Variation (%)
Paired Absolute
Dierence (mg/dL)
Paired Absolute
Relative Dierence
(%)
Number of Paired
Readings
Adults ages 18+
5.7 12.4 8.1 26791
Table 9b: Overall between Sensor Precision (Pediatric; n=137)
Coecient of
Variation (%)
Paired Absolute
Dierence (mg/dL)
Paired Absolute
Relative Dierence
(%)
Number of Paired
Readings
Children ages 4-5
4.8 10.7 6.7 248
Children ages 6-17
5.8 13.0 8.2 10623
212