IMMUNITY
test
IEC 60601
test level
Compliance
Level
Electromagnetic
environment – guidance
Conducted RF
IEC 61000-4-6
6 Vrms
150 kHz to
80 MHz
6 Vrms Portable RF communications
equipment (including peripherals
such as antenna cables and external
antennas) should be used no closer
than 30 cm (12 inches) to any part
of the System, including cables
specified by Abbott Diabetes Care.
Otherwise, degradation of the
performance of the System could
result.
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
See table on
next page
Compliance to
the tested levels
129