EasyManua.ls Logo

Abbott FreeStyle Libre - Control Solution Testing

Abbott FreeStyle Libre
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Control Solution Testing
You should do a control solution test when you are not sure of your test
strip results and want to check that your Reader and test strips are working
properly. You can do a control solution test with a blood glucose or ketone
test strip.
IMPORTANT:
• Control solution results should fall within the control
solution range printed on the test strip instructions for use.
• Do NOT use control solution past the expiry date. Discard
control solution 3 months after opening.
• The control solution range is a target range for control
solution only, not for your blood glucose or ketone results.
• The control solution test does not reect your blood glucose
or ketone levels.
• Use only MediSense glucose and ketone control solution.
• Check that the LOT number printed on the test strip foil
packet and instructions for use match.
• Replace the cap securely on the bottle immediately after use.
• Do NOT add water or other liquid to the control solution.
• Contact Customer Service for information on how to obtain
control solution.
52

Other manuals for Abbott FreeStyle Libre

Related product manuals