EasyManuals Logo

Abbott Precision Xceed Pro User Manual

Abbott Precision Xceed Pro
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
What You See on the Display What You Do Comments
4-4
Questions? Refer to Page 10-10 for Technical Support Instructions.
9. Wait for the monitor to analyze the
sample and display the test result.
The monitor counts down then displays
the test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
10. Note the test result and whether it
falls within the acceptable range.
If required, scan or manually enter the
comment code, and press Enter.
If there is no prompt to enter a
comment code, skip down to step 11.
Control test results can appear as numeric
or PASS/FAIL, depending on how the
test is set up. If numeric, an “X” will appear
in front of a test result that is out-of-range.
The monitor may be enabled to scan or enter
a 1- to 2-digit comment code.
11. You can select one of the following:
Press 1 – Next Level.
Press 2 – Repeat Test.
Press Menu to return to the
Menu Mode menu.
Press On/Off to turn off the monitor.
Remove the test strip from the monitor
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Level (Low, Mid, High)
Comment Code (Number in box)
Test Type (Glucose)
Date and Time of test
Operator ID (o: prompt)
Note that test date and time alternates with
Operator ID.
If 1 is selected, the Scan or Enter Control
Lot screen will appear. Return to step 4 to
perform the Next Level control test.
If 2 is selected, the Scan or Enter Strip
Lot screen will appear. Return to step 5 to
Repeat Test at the same level.
Note: If β-ketone testing is disabled, then following a complete set of Glucose controls,
1-Next Level becomes 1-Exit.
If β-ketone testing is enabled and Glucose QC is run first, then following a complete set of
Glucose controls, 1-Next Level becomes 1-QC Ketone.
If β-ketone testing is enabled and β-ketone QC is run first, then following a complete set of
β-ketone controls, 1-Next Level becomes 1-QC Glucose.
99 mg/dL
Chapter 4 : Control Test

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott Precision Xceed Pro and is the answer not in the manual?

Abbott Precision Xceed Pro Specifications

General IconGeneral
BrandAbbott
ModelPrecision Xceed Pro
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals