61Technical Specications
Technical Features
Continuous Dose Delivery Accuracy:
Continuous dose delivery accuracy testing was performed based upon
IEC 60601-2-24:2012. Testing was performed with distilled water at room
temperature with a 9-mm cannula, 60-cm long Neria Guard infusion set
and with a programmed rate of 0.70 mL/hr: results are shown below.
Mean ow error: -0.14%
Start-up graph
Flow rate
Trumpet curve from minute 472 - 856