EasyManua.ls Logo

AbbVie VYAFUSER - Page 62

AbbVie VYAFUSER
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62 Technical Specications
Technical Features
Bolus Delivery Accuracy:
Bolus delivery accuracy data were generated based upon IEC 60601-2-
24:2012. Testing was performed with distilled water at room temperature
with a 9-mm cannula, 60-cm long Neria Guard infusion set.
Bolus Dose
Volume
Target
Number of
Successive
Bolus
Deliveries
Delivered
Volume
(mL)
Calculated
Mean
Deviation
from Set Value
(mL)
Calculated
Percentage
Deviation from the
Set Value (%)
0.1 mL
(Minimum
bolus volume
setting)
25 2.52755 0.02755
Total volume: 1.1
Max negative: 5.65
Max positive: 4.25
3.0 mL
(Maximum
bolus volume
setting)
3 9.23863 0.23863
Total volume: 2.65
Max negative: 4.23
Max positive: 6.39
Delivery Accuracy Under High Backpressure:
Delivery accuracy data were generated under simulated high
back-pressure conditions (such as from partial occlusions) and using the
time to detect as the test window. This effectively characterizes the mean
ow rate error that could occur under these conditions both including the
initial 30-60 min start-up period, and within the steady-state period to
which the system subsequently converges.
Delivery Accuracy Under High Backpressure
Target Flow Rate
Mean Flow Rate Error
(Including Start-Up)
Mean Flow Rate Error
(Steady-State)
3.0 mL bolus -4.84% --
1.25 mL/hr -10.28% -0.68%
0.70 mL/hr -8.84% -0.94%
0.15 mL/hr -16.08% -4.48%