EasyManua.ls Logo

Access Sensor Technologies UPAS v2 - Interpreting UPAS Status; Shutdown Mode Codes; Conducting a Diagnostic Test (DIAG#)

Access Sensor Technologies UPAS v2
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Access Sensor Technologies UPAS User Guide 55
Duty Cycle: The preset duty cycle (%) is essentially the input / control you
can adjust to extend the battery endurance. Use the UPAS battery
energy management calculator available at www.accsensors.com
to help
determine a duty cycle percentage.
External Battery: If the added weight/complexity are allowable, connect-
ing an external battery pack to the UPAS micro-USB port via cable adds
the capacity of the external pack to the UPAS internal battery capacity.
The Voltaic V15 (https://www.voltaicsystems.com/v15
)and V44
(https://www.voltaicsystems.com/v44
)models in Always-on mode are the
only known packs that reliably provide such extension to the UPAS (other
packs without the Always-on mode may fall asleep due to the nature of
the UPAS charging circuit).
Charging: Connecting the UPAS via USB cable to an AC USB charger
ensures the UPAS can sample indefinitely. This may be an option for
some fixed-location sampling deployments.
Interpreting UPAS Status / Shutdown Mode Codes
The UPAS monitors its status continuously during actively sampling, from
initialization to shutdown. Unless user-disabled via the App, the status indicator
light reports the status in real time, and alerts you if there is a problem before
shutting the UPAS down.
Solid Green indicator light indicates normal sampler function within nominal
parameters.
UPAS interprets its volumetric flow rate to be within the
± 4% of set
point error band (Note that the flow-rate of the UPAS must be periodically
checked vs. a qualified flow rate primary (see “Flow Rate Verification and
Adjustment” on page 24; if necessary to adjust flow rate, see “Adjusting
the Flow Rate using the Flow Offset in the App” on page 45).
Internal battery has more than 15% charge capacity.
UPAS has not exceeded the maximum allowable drive power level.
The internal circuit board is at a temperature of <60
°C.
Flashing Yellow indicator light means the internal battery has been depleted
to below 15%, and the sampler is at risk of shutting down soon.
If it is desired to continue the sample run for much longer the internal
battery will need to be charged soon (see “Charging the UPAS Internal
Battery” on page 20).

Table of Contents