42 Access Sensor Technologies UPAS User Guide
Setting a Start Time for the Next Sample Run
Using the Start With Set Time = Now Option
If starting a sampling run immediately is desired, follow the steps:
1Verify that Select START WITH SET TIME is displayed (or toggle START ON
NEXT POWER-UP to turn it back to START WITH SET TIME).
2Verify that START TIME = NOW.
3 Then when ready to begin sampling, click the START button and OK on the
resulting Pop-up.
The following table explains the Status Indicator Lights as the UPAS initializes:
Active Sampling Sequence if START WITH SET TIME = NOW Option Is Used
Step Sequence Item UPAS Status Indicator Definition
1
Click START to initialize
sampling sequence.
Status light is cyan
After clicking START, the UPAS
disconnects from the mobile
device/APP.
2
Active sampling/pumping
begins. UPAS flow control
not yet set.
Status light turns red
Pumping starts, UPAS adjusts
flow rate.
3
Active sampling/pumping
within flow control
parameters.
Status Light turns green (or is off if
toggled off in App sample window).
UPAS now sampling normally,
with flow control within
parameters.