Cancel “Quick” Standard Bolus
During programming (bolus amount blinks):
• Press the Up or Down key used in Step 1 on prior page to reset the bolus amount
to 0.0 units. The pump will play a melody. If you do not program a new bolus after
5 seconds, your insulin pump will return to RUN mode automatically and beep
three times. No bolus is delivered.
During confi rmation (beeps and vibrations occur), or during the start delay (Standard
bolus symbol blinks):
Press the Up key or Down key. The pump will play a melody. An A8: BOLUS
CANCELLED alert is displayed. Press the Check key twice to confi rm and turn off the
alert. No bolus is delivered.
During bolus delivery (countdown of bolus amount):
Press and hold the Up key or Down key for 3 seconds until you hear a melody. An
A8: BOLUS CANCELLED alert is displayed. Press the Check key twice to confirm
and turn off the alert. Bolus delivery is interrupted.
The bolus amount delivered up to the cancellation can be reviewed in the
BOLUS HISTORY screen.