48
Alerts
Alert
Code Name Meaning Action
A1 Cartridge The cartridge is almost empty • Press the Check key twice to turn
Low (it has only 20 units of U100 off and confi rm the alert.
insulin remaining). • Change the cartridge before
it’s empty.
A2 Battery The battery is low. • Press the Check key twice to turn
Low off and confi rm the alert.
• Change the battery within the
next few hours.
• Put your insulin pump into RUN mode.
A3 Review The battery has been removed • Press the Check key twice to turn
Time and for more than 1 hour and the off and confi rm the alert.
Date time and date need to be • Review the time and date and
re-entered. make any necessary changes.