48
Alerts
Alert
Code Name Meaning Action
A1 Cartridge The cartridge is almost empty Press the Check key twice to turn
Low (it has only 20 units of U100 off and confi rm the alert.
insulin remaining). Change the cartridge before
it’s empty.
A2 Battery The battery is low. Press the Check key twice to turn
Low off and confi rm the alert.
Change the battery within the
next few hours.
Put your insulin pump into RUN mode.
A3 Review The battery has been removed Press the Check key twice to turn
Time and for more than 1 hour and the off and confi rm the alert.
Date time and date need to be Review the time and date and
re-entered. make any necessary changes.