49
Alert
Code Name Meaning Action
A4 Alarm Goes off at the time you have • Press the Check key twice to turn
Clock programmed on the pump. off and confi rm the alert.
• Put your insulin pump into RUN
mode if necessary.
A6 TBR A temporary increase or • Press the Check key twice to turn
Cancelled decrease in the basal rate has off and confi rm the alert.
been cancelled. • Put your insulin pump into RUN
Note: Make sure that you mode if necessary.
intended to cancel the TBR,
and if necessary, program a
new TBR.
A7 TBR A Temporary Basal Rate has • Press the Check key twice to turn
Over ended. off and confi rm the alert.
Note: Your original basal rate
(100%) will resume automatically.
A8 Bolus A bolus has been cancelled • Press the Check key twice to turn
Cancelled during the start delay or after off and confi rm the alert.
delivery has begun. • Put your insulin pump into RUN
Note: Make sure that you mode if necessary.
intended to cancel the bolus,
and if necessary, program a new bolus.