50
Errors
Errors
Code Name Meaning Action
E1 Cartridge The cartridge is empty. Change the cartridge immediately.
Empty • Press the Check key twice to turn
off and confi rm the error.
• Disconnect or remove the infusion
set from your infusion site.
• Change the cartridge and prime the
infusion set.
• Reconnect the infusion set (tubing)
at your infusion site.
•
Put your insulin pump into RUN
mode.
E2 Battery The battery is depleted. Replace the battery immediately.
Depleted • Press the Check key twice to turn
off and confi rm the error.
• Replace the battery.
• Put you insulin pump into RUN mode.