50
Errors
Errors
Code Name Meaning Action
E1 Cartridge The cartridge is empty. Change the cartridge immediately.
Empty Press the Check key twice to turn
off and confi rm the error.
Disconnect or remove the infusion
set from your infusion site.
Change the cartridge and prime the
infusion set.
Reconnect the infusion set (tubing)
at your infusion site.
Put your insulin pump into RUN
mode.
E2 Battery The battery is depleted. Replace the battery immediately.
Depleted Press the Check key twice to turn
off and confi rm the error.
Replace the battery.
Put you insulin pump into RUN mode.