Cancel “Scroll” Standard Bolus continued
During bolus delivery (countdown of bolus amount):
Press and hold the Up key or Down key for 3 seconds until the pump plays a
melody. An alert A8: BOLUS CANCELLED occurs. Press the Check key twice to
confi rm and turn off the alert. Bolus is interrupted.
The bolus amount delivered up to the cancellation can be reviewed in the
BOLUS HISTORY screen.