33
Program MultiWave Bolus continued
TIP:
As you increase or decrease the TOTAL BOLUS AMOUNT, the suggested immediate
bolus amount increases or decreases proportionally.
Step 4: Press the Menu key to move to the IMMEDIATE BOLUS screen.
Step 5:
Press or scroll the Up key or Down key to select the immediate bolus amount.
Step 6: Press the Menu key to move to the BOLUS DURATION screen.
Step 7: Press or scroll the Up key or Down key to select the extended bolus
duration.
TIP: Press the Menu key to loop between programming the TOTAL BOLUS AMOUNT,
IMMEDIATE BOLUS AMOUNT, and EXTENDED BOLUS DURATION, if necessary.
Step 8: Press the Check key to confi rm both bolus amounts and the bolus
duration.