Program Temporary Basal Rate (TBR) continued
Step 1: Put your insulin pump in Run mode.
Step 2: Press the Menu key to move to the TEMPORARY BASAL RATE (TBR) screen.
Step 3: Press the Check key to select.
Step 4: Press or scroll the Up key to increase or the Down key to decrease the basal
rate in increments of 10% from 0% to 200%.
Step 5: Press the Menu key to move to the TBR DURATION screen.
Step 6: Press or scroll the Up key or Down key to correct and set the TBR DURATION.
Step 7: Press the Check key to save and exit. Your Temporary Basal Rate will activate
immediately.
A TBR is displayed, which will remain for the duration you set in the TBR duration unless
you cancel this setting. When the TBR has ended, an A7: TBR OVER alert occurs. Press
the CHECK key twice to confi rm and turn off the alert.