174 C
"HORZ_ONE_RESET_TOUT
Timeout of primary collimator
horizontal axis reset"
U/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check motor and its cable;
check optoswitch and its cable;
check/replace AUX CBCT board."
175 C
"HORZ_ONE_PT_PT_GOTO
The point-to-point positioning of
horizontal axis of primary collimator
is failed"
A
Reset error from workstation; if error remains restart trium and
workstation.
177 C
"XRAYS_EXAM_DO_BREAK
WARNING: operator has intentionally
stopped the Xray emission releasing
XRay button"
U
Wait the full exam procedure is completed before releasing
XRay button: message onworkstation “exposure button can be
released”