Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
ENGLISH
163 C
"AUX_CBCT_LINK_ERROR
CAN communication error between
Mainboard and AUX CBCT board"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check CAN cables between Maiboard and AUX CBCT;
check/replace Mainboard;
check/replace AUX CBCT board."
164 B
"AUX_CEPH_LINK_ERROR
CAN communication error between
Mainboard and AUX CEPH board"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check CAN cables between Maiboard and AUX CEPH;
check/replace Mainboard;
check/replace AUX CEPH board."
165 C
"PRM_COLLIM_DO_BREAK
WARNING: operator has intentionally
stopped a movement releasing XRay
button"
U
Wait the full exam procedure is completed before releasing
XRay button: message on workstation “exposure button can be
released”
166 C
"PRM_COLLIM_CNTL_FSM
Unexpected primary collimator FSM
condition"
A
Reset error from workstation; if error remains restart trium and
workstation.
167 B
"VERT_INF_RESET_RAMP
Unexpected FSM condition during
reset of primary collimator vertical-inf
axis"
A
Reset error from workstation; if error remains restart trium and
workstation.
168 C
"VERT_INF_RESET_TOUT
Timeout of primary collimator vertical-
inf axis reset"
U/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check motor and its cable;
check optoswitch and its cable;
check/replace AUX CBCT board."
169 C
"VERT_INF_PT_PT_GOTO
The point-to-point positioning of
vertical-inf axis of primary collimator
is failed"
A
Reset error from workstation; if error remains restart trium and
workstation.
170 B
"VERT_SUP_RESET_RAMP
Unexpected FSM condition during
reset of primary collimator vertical-
sup axis"
A
Reset error from workstation; if error remains restart trium and
workstation.
171 C
"VERT_SUP_RESET_TOUT
Timeout of primary collimator vertical-
sup axis reset"
U/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check motor and its cable;
check optoswitch and its cable;
check/replace AUX CBCT board."
172 C
"VERT_SUP_PT_PT_GOTO
The point-to-point positioning of
vertical-sup axis of primary collimator
is failed"
A
Reset error from workstation; if error remains restart trium and
workstation.
173 B
"HORZ_ONE_RESET_RAMP
Unexpected FSM condition during
reset of primary collimator horizontal
axis"
A
Reset error from workstation; if error remains restart trium and
workstation.