152 C
"COL_DCFI_SFTW_RESET
Hanning board error code 133 - Reset
by SW"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
153 C
"COL_DCFI_E2PROM_CRC
Hanning board error code 140 -
Parameter memory CRC error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
154 C
"COL_DCFI_E2PROM_TAB
Hanning board error code 141 -
Parameter memory table has an
error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
155 C
"COL_DCFI_DFLT_TABLE
Hanning board error code 142 -
Parameter memory factory default
table has an error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
156 C
"COLUMN_POS_WDG_TOUT
Column motion stopped due to CAN
communication timeout between
Mainbord and Power"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check CAN cable between Maiboard and Power;
check/replace Mainboard;
check/replace Power board."
157 C
"COLUMN_POS_END_LIMI
Column motion stopped because of
the achievement of mechanical limit
switches"
U/T
"Reset error from workstation.
Move the column in the opposite direction.
The equipment can be operated, but the occurrence of
this error means that the column potentiometer is out of
calibration:
Technical service is required:
carry out column potentiometer calibration"
158 B
"COLUMN_POS_RUN_LIMI
Column motion stopped because of
the achievement of calibrated upper
or lower position"
U Only a warning: operator can move in the opposite direction.
159 C
"COLUMN_POT_REF_LIMI
Reference value for column
potentiometer is out of range"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check column potentiometer and its cable;
check/replace Power board."
160 C
"COLUMN_POT_POS_LIMI
Read value of column potentiometer
is out of range"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check column potentiometer and its cable;
check/replace Power board."
161 C
"COLUMN_POT_POS_BLCK
Column motion stopped because of
the column potentiometer is blocked"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check column potentiometer and its cable;
check column motor and its cable;
check/replace column (Hanning) board;
check/replace Power board."