Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
ENGLISH
139 D
"EMERGENCY_STOP_DONE
Trium has rebooted after releasing
the emergency stop button"
U Reset error from workstation.
140 C
"COL_DCFI_LW_VOLTAGE
Hanning board error code 1 - Under-
voltage"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
141 C
"COL_DCFI_HI_VOLTAGE
Hanning board error code 2 - Over-
voltage"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
142 C
"COL_DCFI_HI_TEMPRAT
Hanning board error code 5 - Over-
heated motor or inverter"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
143 C
"COL_DCFI_HI_CURRENT
Hanning board error code 9 - Inverter
over-current / peak current error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
144 C
"COL_DCFI_SHORT_CIRC
Hanning board error code 13 - Short-
circuit – shut down"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
145 C
"COL_DCFI_IS_NOT_ENA
Hanning board error code 16 - Not
enabled"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
146 C
"COL_DCFI_TXCMD_TOUT
Hanning board error code 17 - Time-
out – digital interface"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
147 C
"COL_DCFI_ON_DIR_ERR
Hanning board error code 18 - Start
attempt with directional error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
148 C
"COL_DCFI_TIMING_ERR
Hanning board error code 128 -
Internal timing error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
149 C
"COL_DCFI_SYSTEM_ERR
Hanning board error code 129 -
System error"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
150 C
"COL_DCFI_WDOG_RESET
Hanning board error code 131 - Reset
by Watch Dog"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."
151 C
"COL_DCFI_VDIP_RESET
Hanning board error code 132 - Reset
by voltage dip (brown-out)"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace column (Hanning) board."