49
ENGLISH
Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
128 C
"IS_OVER_LOAD_ON_POS
Inverter board error:
GENERAL STATE ERROR, bit 37 -
OVERLOAD ON POSITIVE HV STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
129 C
"IS_OVER_LOAD_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 38 -
OVERLOAD ON NEGATIVE HV STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
130 C
"NO_FBACK_ON_POS_ERR
Inverter board error:
GENERAL STATE ERROR, bit 40 - NO
FEEDBACK ON POSITIVE HV STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
131 C
"NO_FBACK_ON_NEG_ERR
Inverter board error:
GENERAL STATE ERROR, bit 41 - NO
FEEDBACK ON NEGATIVE HV STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
133 C
"POWERDEV_LINK_ERROR
CAN communication error between
Mainboard and Power
"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check CAN cable between Maiboard and Power;
check/replace Mainboard;
check/replace Power board."
134 C
"COL_DCFI_LINK_ERROR
CAN communication error between
Power and Hanning board (column
board)"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check CAN cable between Power and column (Hanning) board;
check/replace column (Hanning) board;
check/replace Power board."
135 C
"PCF_VOLT_IS_UPP_LIM
PFC over voltage detected"
A/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace Power board."
136 C
"PCF_VOLT_IS_LOW_LIM
PFC under voltage detected"
A/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace Power board."
137 C
"PCF_TEMP_IS_UPP_LIM
PFC over temperature detected"
A/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace Power board."
138 C
"PCF_TEMP_IS_LOW_LIM
PFC under temperature detected"
A/T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check/replace Power board."