48
Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
ENGLISH
119 C
"EXP_TIME_IS_LOW_LIM
Inverter board error:
GENERAL STATE ERROR, bit 28 -
EXPOSURE TIME TOO SHORT"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
120 C
"EXP_TIME_IS_UPP_LIM
Inverter board error:
GENERAL STATE ERROR, bit 29 -
EXPOSURE TIME TOO LONG"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
121 C
"EXP_NUM_PLS_LOW_LIM
Inverter board error:
GENERAL STATE ERROR, bit 30 - TOO
FEW EXPOSURES"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
122 C
"EXP_NUM_PLS_UPP_LIM
Inverter board error:
GENERAL STATE ERROR, bit 31 - TOO
MANY EXPOSURES"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
123 C
"TUBEH_IS_NOT_DETECT
Inverter board error:
GENERAL STATE ERROR, bit 32 - RX
TUBE UNIT NOT CONNECTED OR
TEMPERATURE SENSOR FAULT"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
124 C
"TUBEH_OVER_TEMP_ERR
Inverter board error:
GENERAL STATE ERROR, bit 33 - RX
TUBE UNIT OVER TEMPERATURE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
125 C
"TUBEH_SENS_TEMP_ERR
Inverter board error:
GENERAL STATE ERROR, bit 34 - RX
TUBE UNIT TEMPERATURE SENSOR
FAULT"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
126 C
"NO_CURR_GENR_ON_POS
Inverter board error:
GENERAL STATE ERROR, bit 35 - NO
CURRENT GENERATED BY POSITIVE HV
STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
127 C
"NO_CURR_GENR_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 36 - NO
CURRENT GENERATED BY NEGATIVE
HV STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."