47
ENGLISH
Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
110 C
"VOLT_UPP_LIM_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 15 -
VOLTAGE TOO HIGH ON NEGATIVE
STAGE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
111 C
"HI_VOLTS_ARC_ON_POS
Inverter board error:
GENERAL STATE ERROR, bit 16 - ARC
DETECTED ON POSITIVE HV"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
112 C
"HI_VOLTS_ARC_ON_NEG
Inverter board error:
GENERAL STATE ERROR, bit 17 - ARC
DETECTED ON NEGATIVE HV"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
113 C
"ANOD_CURRENT_ABSENT
Inverter board error:
GENERAL STATE ERROR, bit 19 -
ANODIC CURRENT ABSENT"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
114 C
"ANOD_CURR_LOW_LIMIT
Inverter board error:
GENERAL STATE ERROR, bit 20 -
ANODIC CURRENT TOO LOW"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
115 C
"ANOD_CURR_UPP_LIMIT
Inverter board error:
GENERAL STATE ERROR, bit 21 -
ANODIC CURRENT TOO HIGH"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
116 C
"HEAT_SINK_OVER_TEMP
Inverter board error:
GENERAL STATE ERROR, bit 23 - HEAT
SINK OVER TEMPERATURE"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
117 C
"FILM_CURR_LOW_LIMIT
Inverter board error:
GENERAL STATE ERROR, bit 24 -
FILAMENT CURRENT TOO LOW"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."
118 C
"FILM_CURR_UPP_LIMIT
Inverter board error:
GENERAL STATE ERROR, bit 25 -
FILAMENT CURRENT TOO HIGH"
T
"Reset error from workstation; if error remains restart trium
and workstation.
If error remains technical service is required:
check cables between Inverter board and tubehead;
check/replace tubehead;
check/replace Inverter board."