Either place 10 parts on the platform for determining the average piece
weight or use a different number of parts. For example, place 20 parts on
the platform, press [CE] to clear the last values and then enter the value
20 using the numeric keypad.
Press [Cnt] to weigh the samples and determine an average piece
weight.
If the parts are too light to measure accurately, the count may become
faulty. It is suggested that the samples to be weighed should each weigh
more than the resolution of the scale.
After the sample has been weighed the scale will count any other parts
added by applying the average piece weight to the weight of the parts to
be counted.
The [Tare] key works normally during this time, so it is possible to tare
the display with a container on the platform or to enter a preset tare
value as described in section 10.2.2.
During parts counting the display can be changed to show the net
weight, unit weight and number of parts by each time pressing the
[Func] key.