What to do if Advanced Instruments Measuring Instruments shows error message “No Plateau”?
- EErin PerryJul 29, 2025
If your Advanced Instruments Measuring Instruments displays the error message “No Plateau,” it means the instrument couldn't detect a freeze plateau and couldn't provide a result. You can retest the sample or run Clinitrol 290 Reference Solution. Also, confirm that you are using proper technique and that the sampler is in good condition. A sample probe problem is also possible.