information in memory, or if a probe has been replaced, the results
displayed may be far from the nominal value of the standards used.
If the displayed values repeat consistently, the calibration will auto-
matically adjust when the calibration sequence is complete.
• The calibration procedure may be terminated at any time by press-
ing [EXIT]. The instrument will display "Calibration Canceled",
and beep twice. The previous calibration will be retained. The user
will be prompted again to insert the appropriate calibration stan-
dard.
• The exact calibration procedure will depend upon whether or not
the optional 2000 mOsm Calibration point has been turned ON in
the Setup Menu at the time of calibration. The instrument is
shipped from the factory with the 2000 mOsm calibration point
turned ON. Please pay close attention to the prompts shown on the
instrument display.
21
Calibration