ADC Compact
Type 5145 / 100/101/140
Type 5145 / 200/201/240
Work Instruction for order no. SN Cycles
The maintenance has to be carried out:
· Every 6 months or every 15 000 cycles
Maintenance must be carried out according to the maintenance instructions,
see section 12 of your Service Documentation.
Maintenance Step by Step Done
Diagnostics
· Ask customer about problems
· Read out the info counters, analyse them and clear
them afterwards
· Check visually the overall condition
o
o
o
Inside
· Vacuum the inside and wipe it
o
Cassette Unit
· Clean opener mechanism
· Check roller for visible wear and exchange if necessary
· Exchange belt once a year
o
o
o
Power Unit
· Check electrical function of safety switch
· Exchange air filter
o
o
VME-Rack
· Check electrical function of the VME-Rack fan
· Remove dust and dirt
o
o
Scan Unit
· Clean scan-rollers with ADC Cleaner
(do not remove!)
· Check discharge brush and clean or replace it if
necessary
o
o