Maintenance Checklist
DD+DIS014.02E
Maintenence Checklist ADC Compact Edition 2, Revision 8
Page 2 (Type 5145 / 100/101/140/200/201/240)
Maintenance Step by Step Done
Transport Units
· Check suction cups position, clean them with ADC
Cleaner and exchange suction cups once a year
· Check toothed belt and pulley at standard robot
o
o
Erasure Unit
· Remove dust from reflector, input and output opening of
air stream, KG2 filter, outer front panel
· Check the KG2 filter
· Exchange lamps every maintenance
o
o
o
Cassettes
· Check the following items of the cassette:
outside condition, hinges, locking, opening leaf springs,
aluminum label
· Attach missing aluminum labels
o
o
Image Plates
· Check if there are scratches on the surface
· Check if edges are loose
o
o
Checking the Image Quality Done
General
· Check the last 20 to 40 images on the VIPS
· Carry out test cycles, four with each format
· Print flat field sample provided by the digitizer
· Expose a flat field
· Evaluate the flat field and carry out respective actions
o
o
o
o
o
Completion of the maintenance Done
General
· Confirm the maintenance
· Make a backup of the system on floppy
· Inform the customer about maintenance and which
repairs are necessary
o
o
o