10
2241A EN 20040210
You can hurt your fingers if they are caught between the ADC Cassette and the
edge of the input slot. Insert the cassette in the input buffer as described in
‘Reading an image plate’ on page 28. At all times, keep your fingers clear of the
input slot. As soon as the CR 75.0 takes in the cassette, release it.
TÜV safety issues
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards (e.g. IEC 950 for data
processing equipment and IEC 601-1 for medical equipment). Furthermore
all configurations shall comply with the valid version of the system standard
IEC 601-1-1. Everybody who connects additional equipment to the signal
input part or signal output part configures a medical system, and is therefore
responsible that the system complies with the requirements of the valid
version of the system standard IEC 601-1-1. If in doubt, consult your local
service organization.
Power On
Power Off
Note that the power cord has to be disconnected from the wall
outlet in order to disconnect the unit entirely from the mains.
Precautions for use in USA only:
Make sure that the circuit is single-phase center-tapped, if the
Digitizer is connected to a 240 V/60 Hz source instead of a 120 V/
60 Hz source.