Safety warning, indicating that the CR 75.0 manuals should be
consulted before making any connections to other equipment. The
use of accessory equipment not complying with the equivalent
safety requirements of this Digitizer may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice
of accessory equipment shall include:
• Use of the accessory equipment in the patient vicinity,
• Evidence that the safety certification of the accessory equipment
has been performed in accordance with the appropriate IEC 601-
1 and IEC 601-1-1 harmonized national standard.
In addition all configurations must comply with the medical
electrical systems standard IEC 601-1-1. The party that makes the
connections acts as system configurator and is responsible for
complying with the systems standard.
If required contact your local service organization.
In order to reduce the risk of electric shock, do not remove any
covers.
Caution hot:
Keep hands clear from the erasure unit.
Type B equipment:
Indicates that the CR 75.0 complies with the limits for type B
equipment.
Supplementary protective earth connector:
Provides a connection between the CR 75.0 and the potential
equalization busbar of the electrical system as found in medical
environments. This plug should never be unplugged before the
power is turned off and the power plug has been removed.
Intergrounding connector:
Provides a connection between the Digitizer and other equipment
which might exhibit minor ground potential differences. These
differences may degrade the quality of communication between
different equipment. Never remove connections to this terminal.
Protective earth (ground):
Provides a connection between the Digitizer and the protective
earth of the mains. Do not remove this connection, because this
will have a negative influence on the leakage current.