Symptom Cause Remedy
Check if the wireless network
communication is stable.
All status indica-
tors are flashing.
A hardware error has
occurred.
Turn off the DR Detector and
turn it on again.
Two status indi-
cators are flash-
ing and the third
is flashing slow-
ly.
A fully charged
battery is con-
sumed quickly.
The battery capacity de-
creases.
The DR Detector battery can
deteriorate because of its char-
acteristics and structure. For
purchase of consumables, con-
tact your sales representative or
local dealer.
The battery was charg-
ed or used in low tem-
peratures.
In low temperatures the battery
capacity decreases. Use a bat-
tery charged in normal temper-
atures.
The battery bay
is unusually hot.
The battery is malfunc-
tioning.
Stop using the battery and con-
sult your sales representative or
local dealer.
104 | DX-D 45C, DX-D 45G | Problem solving