Safety directions for the System Control Unit
WARNING:
Do not block the ventilation ports to prevent overheating.
Overheating can cause system malfunction and damages.
WARNING:
Ensure continuous power supply to the system, with voltage and
current according to the product specifications. If power failures
are frequent, an uninterrupted power supply (UPS) should be
installed to avoid loss of data.
CAUTION:
The System Control Unit and the X-ray generator must be
grounded to a common protective earth. Always connect the
three-core power cord plug to a grounded AC power outlet.
DX-D 45C, DX-D 45G | Introduction to the DR Detector | 71
0292B EN 20190401 1627