DX-D 45C, DX-D 45G
Commercial name DX-D 45C, DX-D 45G
Electrical connection DR Detector
Rated power supply
(powered by battery pack)
DC +24 V, Max. 0.8 A
Power consumption
max. 19.2 W
Wireless connection IEEE 802.11n (2.4 GHz/5 GHz)
Wireless signal range (in an open
space)
maximum 8 m
Environmental conditions (during normal operation)
Room temperature between +10 °C and +35 °C
Humidity (non condensing) between 30% and 85% RH
(non-condensing)
Atmospheric pressure between 700 hPa and 1060 hPa
Environmental conditions (during storage and transport)
Temperature (ambient) between -15 °C and +55 °C
Humidity (non condensing) between 10% and 90%
(non-condensing)
Atmospheric pressure between 500 and 1060 hPa
Warming-up time
30 minutes
Dimensions
Dimensions
width x length x height
approx. 287 x 350 x 15.0 mm
Weight (incl. battery)
2.2 kg
Maximum load 100 kg on an area of 40 mm in diameter
Maximum total load 150 kg over the whole detector surface
106 | DX-D 45C, DX-D 45G | Technical Data
0292B EN 20190401 1627