Note: This product may malfunction due to electromagnetic waves
caused by portable personal telephones, transceivers, radio-
controlled toys, etc. Be sure to avoid having objects such as these,
which affect this product, brought near the product.
CAUTION:
Sudden heating of the room in cold areas will cause
condensation to form on the equipment. In this case, wait until
the condensation evaporates before use. If the equipment is
used while condensation is formed on it, problems may occur.
When an air-conditioner is used, be sure to raise/lower the
temperature gradually so that a difference in temperature in the
room and in the equipment does not occur, to prevent
condensation.
WARNING:
Do not use non-medical equipment in the patient’s vicinity.
Figure 8: Patient’s vicinity