79
2321 E EN 20130722
DX-G/DX-M
Power consumption (typical values)
220V - 240V / 50-60Hz configuration
•Standby
87 W
•Average continuous scanning
237 W
•During scanning
590 W max.
100V - 120V / 50-60Hz configuration
•Standby
92 W
•Average continuous scanning
245 W
•During scanning
621 W max.
Environmental conditions (during operation)
In line with IEC721-3-2: 2M2 and 2K2 with following restrictions:
Room temperature Agfa CR HM5.x Mammo: between
+20 °C and +30 °C (68 - 86°F)
Other plates and cassettes: between
+15 °C and +30 °C (59 - 86°F)
Maximum temperature change rate max. 0.5 °C/min (0.9°F/min)
Relative humidity between 15 % and 75 %
(non-condensing)
Light no exposure to direct sunlight, max. 2500 Lux
Environmental conditions (during storage)
Room temperature between -25 °C and 55 °C
Relative humidity between 10 % and 95 %
Barometric pressure between 70 kPa and 106 kPa
Environmental conditions (during transport)
Room temperature between -25 °C and 55 °C
Relative humidity maximum 85 %
Barometric pressure minimum 70 kPa
Environmental conditions for mobile installation (during transport)
In line with IEC721-3-5: 5M1 and 5K1 with following restrictions:
Room temperature between 10 °C and 40 °C