EasyManua.ls Logo

AGFA DX-M - Page 80

AGFA DX-M
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80
2321 E EN 20130722
DX-G/DX-M
Environmental conditions for mobile installation (during operation)
In line with IEC721-3-3: 3M1 and 3K2 with following restrictions:
Room temperature Agfa CR HM5.x Mammo: between
+20 °C and +30 °C (68 - 86°F)
Other plates and cassettes: between
+15 °C and +30 °C (59 - 86°F)
Relative humidity between 15 % and 75 %
(non-condensing)
Vibration 40 - 200 Hz; 1 m/s²; sinusoidal vibration
Barometric pressure/altitude
Highest pressure 106 kPa (at sea level)
Lowest pressure 70 kPa (3000 m above sea level)
Warming-up time
Cold start 3 min.
Physical emissions
Noise emission (sound power level according to ISO 7779)
During scanning max. 65 dB(A)
Standby max. 45 dB(A)
Heat emission per hour (typical values)
During continuous operation (with
standard erasure time)
245 Wh / 836 BTU
Standby 92Wh / 314BTU
Plates and cassettes
Compatible cassette/IP types DX-G
Agfa CR HD5.x General/FLFS/AEC.
Agfa CR MD4.xR General/FLFS.
Compatible cassette/IP types DX-M
Agfa CR HD5.x General/FLFS/AEC.
Agfa CR MD4.xR General/FLFS.
Agfa CR HM5.x Mammo.
Agfa CR MM3.xR Mammo.

Table of Contents