Artifact in DR Detector images
Details An artifact is visible in the images produced by a DR
Detector.
Cause Exposure conditions have changed significantly since
latest calibration.
Brief Solution Perform calibration of the DR Detector.
For details, refer to the DR Detector Calibration Key
User Manual (document 0134).
XD 14, XD14 | Problem solving | 115
0430A EN 20210426 1355