Automatic exposure detection
Depending on the configuration, the DR detector detects X-ray exposure to
automatically perform the image acquisition.
Before performing the exposure, the DR detector must be ready. Check the
status of the DR detector in the DR Detector Switch.
WARNING:
The sensor for automatic exposure detection must be in the
exposed area. Positioning the sensor for automatic exposure
detection outside the exposed area can cause failure to trigger
the image acquisition.
WARNING:
Very short exposure time can cause failure to trigger the image
acquisition. Use an exposure time of at least 3 ms.
WARNING:
Specific exposure conditions (use of grid, thickness of the
exposed object) can cause failure to trigger the image
acquisition or horizontal artifacts in the acquired image.
Related Links
Positioning the XD 14, XD
14 on page 84
94 | XD 14, XD14 | Getting started