Daily inspection
WARNING:
For safety reasons, be sure to turn OFF the power to each piece
of equipment before performing the following. Otherwise, an
electric shock may result.
1. Ensure that cables are not damaged and cable jackets are not torn.
2. Ensure that the power cord plugs are securely connected to both the
equipment AC inlet and the AC outlet.
3. Ensure that there are no loose screws or breaks.
Turn on the power. Start the MUSICA Acquisition Workstation and perform a
test exposure.
Half-yearly inspection
To indicate when the half-yearly calibration is due, a message is displayed on
the MUSICA Acquisition Workstation.
Perform calibration half-yearly or when exposure conditions have changed
significantly. For details, refer to the DR Detector Calibration Key User
Manual (0134).
XD 14, XD14 | Introduction to the DR Detector | 59
0430A EN 20210426 1355