40
Chapter 3 Front-Panel Menu Operation
To Read the Calibration Information
3
To Read the Calibration Information
You can use the instrument’s calibration memory to read the calibration
count and calibration message.
Calibration Count You can query the instrument to determine how
many calibrations have been performed. Note that your instrument was
calibrated before it left the factory. When you receive your instrument,
read the count to determine its initial value. The count value increments
by one for each calibration point, and a complete calibration may
increase the value by many counts.
Calibration Message The instrument allows you to store one message
in calibration memory. For example, you can store such information as
the date when the last calibration was performed, the date when the
next calibration is due, the instrument’s serial number, or even the name
and phone number of the person to contact for a new calibration.
1 Select the Cal Info interface.
Press and then select the Cal Info softkey from the “Test/Cal” menu.
The first line in the display shows the calibration count.
The second line shows the calibration message.
The last line indicates the current version of the firmware.
The calibration information will time-out and disappear after a few seconds.
Select the Cal Info softkey to show the information again.
2 Exit the menu.
Press the DONE softkey.
You can record a calibration message only from the remote interface
and only when the instrument is unsecured.
You can read the message from either the front-panel or over the remote
interface. You can read the calibration message whether the instrument
is secured or unsecured.