EasyManua.ls Logo

Agilent Technologies 33250A - Internal ADC Adjustment

Agilent Technologies 33250A
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
75
Chapter 4 Calibration Procedures
Internal ADC Adjustment
4
4
Internal ADC Adjustment
The function generator stores calibration constants related to the gain
and offset of the internal ADC. Setup 6 must always be performed
before any other adjustments are attempted. The internal ADC is then
used as a source for the calibration constants generated in setup 7.
1 Make the connections as shown below.
2 Set the DMM to display 5
½ digits and measure the dc value.
Record the measurement.
3 Enter the following setup and use the numeric keypad or knob to enter
the measured value of the dc source.
* Constants are stored after completing this setup.
4 Disconnect all cables from the rear panel Modulation In connector.
Nominal Signal
Setup DC level
6* ~1 Vdc ±10% Calibrates the internal ADC.
Modulation In

Table of Contents

Other manuals for Agilent Technologies 33250A

Related product manuals