Chapter 4 Calibration Procedures
Agilent Technologies Calibration Services
4
4
Closed-Case Electronic Calibration The instrument features closed-
case electronic calibration. No internal mechanical adjustments are
required. The instrument calculates correction factors based upon the
input reference value you set. The new correction factors are stored in
non-volatile memory until the next calibration adjustment is performed.
Non-volatile EEPROM calibration memory does not change when power
has been off or after a remote interface reset.
Agilent Technologies Calibration Services
When your instrument is due for calibration, contact your local
Agilent Technologies Service Center for a low-cost recalibration.
The Agilent 33250A is supported on automated calibration systems
which allow Agilent to provide this service at competitive prices.