EasyManua.ls Logo

Agilent Technologies 33250A - Page 56

Agilent Technologies 33250A
211 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
Chapter 4 Calibration Procedures
Performance Verification Tests
4
Self-Test
A brief power-on self-test occurs automatically whenever you turn on the
instrument. This limited test assures that the instrument is operational.
To perform a complete self-test:
1 Press on the front panel.
2 Select the Self Test softkey from the “Test/Cal” menu.
A complete description of the self-tests can be found in chapter 6.
The instrument will automatically perform the complete self-test
procedure when you release the key. The self-test will complete in
approximately 30 seconds.
If the self-test is successful, “Self Test Pass” is displayed on the
front panel.
If the self-test fails, “Self Test Fail” and an error number are displayed.
If repair is required, see chapter 6, “Service,” for further details.

Table of Contents

Other manuals for Agilent Technologies 33250A

Related product manuals