Chapter 4 Calibration Procedures
Performance Verification Tests
4
Self-Test
A brief power-on self-test occurs automatically whenever you turn on the
instrument. This limited test assures that the instrument is operational.
To perform a complete self-test:
1 Press on the front panel.
2 Select the Self Test softkey from the “Test/Cal” menu.
A complete description of the self-tests can be found in chapter 6.
The instrument will automatically perform the complete self-test
procedure when you release the key. The self-test will complete in
approximately 30 seconds.
If the self-test is successful, “Self Test Pass” is displayed on the
front panel.
If the self-test fails, “Self Test Fail” and an error number are displayed.
If repair is required, see chapter 6, “Service,” for further details.