84
Chapter 4 Calibration Procedures
+10 dB Range Flatness Adjustments
4
* Constants are stored after completing this setup.
3 Using the numeric keypad or knob, adjust the displayed amplitude at
each setup to match the measured amplitude (in dBm).
Select ENTER VALUE.
4 After performing setup 79:
a. If your calibration procedures require you to verify the adjustment just
made, exit the calibration menu and perform “+10 dB Range Flatness
Verification”, on page 65.
b. If you are making all the adjustments and then verifying the instrument’s
performance, continue with the next procedure in this chapter.
Nominal Signal
Setup Frequency Amplitude
66* 100 kHz 0.9 Vrms 12 dBm Power Meter Reference for +10 dB Range
67* 200 kHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
68* 500 kHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
69* 1.5 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
70* 5 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
71* 10.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
72* 25.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
73* 40.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
74* 50.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
75* 60.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
76* 65.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
77* 70.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
78* 75.1 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter
79* 79.9 MHz 0.9 Vrms 12 dBm Flatness for +10 dB, Elliptical Filter