8. Select parameter “11” in the parameter menu and enter the calculated value in the
following manner:
”-80” and <Enter> (for the above mentioned example, if the value of the controller is
higher than the actual value).
”80” and<Enter> (for the above mentioned example, if the value of the controller is
lower than the actual value)
9. Press <Arrow right> to return to the diagnostic menu and leave it with <Enter>.
10. Change the validation mode (ValMode) to the normal measuring mode (Operate) in
the General Configuration Menu (see above and Manual 5.6).