General Maintenance 4
5975/5977 Series MSD Operation Manual 105
To Clean the EI Ion Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (8660-0791)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the ion source. See “To Disassemble the Standard or Inert EI
Ion Source” on page 99 or “To Disassemble the Extractor EI Ion Source” on
page 102.
2 Collect the following parts to be cleaned for a standard or inert EI ion
source: (See Figure 15 on page 107.)
Repeller
Interface socket
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens