EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>5975 MSD

Agilent Technologies 5975 MSD User Manual

Agilent Technologies 5975 MSD
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #72 background imageLoading...
Page #72 background image
72 5975/5977 Series MSD Operation Manual
3 Operating in Electron Ionization (EI) Mode
To Calibrate Column Flow Linear Velocity
Capillary columns must be calibrated prior to use with the MS.
Procedure
1 Set Data Acquisition for splitless manual injection and set up a real time
plot to monitor m/z 28.
2 Press [Prep Run] on the GC keypad.
3 Inject 1 µL of air into the GC inlet and press [Start Run]
4 Wait until a peak elutes at m/z 28. Note the retention time.
5 In OpenLAB, select Method layouts, and click Agilent GC > Configuration >
Columns.
6 Select your installed column from the table.
7 Click the Calibrate button to display the Calibrate Column dialog.
8 Click the Calc Length button in the If unretained peak holdup time is known
section to display the Calculate Column Length dialog.

Table of Contents

Other manuals for Agilent Technologies 5975 MSD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 MSD and is the answer not in the manual?

Agilent Technologies 5975 MSD Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 MSD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals