Operating in Electron Ionization (EI) Mode 3
5975/5977 Series MSD Operation Manual 53
6 Confirm that BootP is set to OFF. If your LAN assigns IP addresses using a
BootP server click ON and skip the next step.
7 To update the MSD IP address, Gateway IPA, and SubNet Mask enter the new
values. Before clicking submit, you can recover the previous settings by
clicking Return to Main Menu and then returning to here.
8 Click Submit to upload this net configuration to the MSD.
A dialog opens to confirm that the network configuration process has
started.
9 Click OK to close the dialog and wait for the prompt requesting to Manually
reboot of the MSD/SmartCard to activate the new Settings.
10 Use the MSD start/stop button to reboot the MSD SmartCard.
Figure 8 WUI Edit NetConfig